Light source: Tungsten filament lamp
Acc. to standard methods: US-EPA 180.1
Modes: Nephelometric (scattering 90°)
Display: Backlit graphics display 160x104
Keypad: Foil keypad with tactile feedback alphanumeric entry option
Range: 0-1100 NTU
Units: NTU
Resolution (Measurement resolution max. 0,02 NTU due to natural scattering of water; digital resolution >0,02 if optional): 0,01 for 0,00 -9,99, 0,1 for 10 – 99,90, 1 for 100 - 1100
Accuracy NTU/FNU: 0.01 NTU or +/- 2% of finding; 500-1100 NTU: +/- 3% of finding
Repeatability: < 1% of finding
Reading modes: Findings with Intelligent Reproducibility and Plausibility Check (IRPC), rapid settlement samples supported by quick response time and IRPC
Response time: 7 sec
Calibration – options: Automatic 3-P Kal, (Default), 2-5 P user calibration, QUICKCal
Calibration protocol and interval setting: Yes/yes
AQS-Support: Kal.-protokoll, AQS-Flag, Kal.-Flag
Data storage: 2500 data sets with Cal protocol, AQA flag, GLP comliant data management with LIMS-connection
Sample Identification: Yes - alphanumeric
Air Purge: No
Firmware-Update: Yes via USB
Interface: RS 232, USB, Printer via PC or RS232
Storage condition instrument: - 25 °C ... + 65 °C (13-149 °F)
Operating Temperature Range: +5 °C ... +55 °C (41-131 F)
With power plug connected: +5 °C ... +40 °C (41-104 F)
Data management: GLP-compliant incl. PC software: USB, RS232 Turb® Data
Dimensions (H x W x D): ca. 290 x 190 x 80 mm
Weight: 1.1 kg
Calibration Standards: Longterm stable Cal.Kits for standard 3-P calibration; 0,02-10-1000 FNU/NTU; Polymer AMCO® Clear standards: longterm stability, higly precise, optimized to optics, production accuracy ± 1%, non-hazardous
Vials, Sample Volume: Vials 28 mm diameter, volume min. 15 ml; Borosilicate, cap: phenolic resin, PTFE coated sealing gasket.; NO SILICON OIL to cover scratches is required for the measurement procedure!
Sample conditions: Sample temperature < 70°C (158°F)
Power supply: Wide range power supply with plugs for Euro, US, UK and Australia
Certificates: CE
Delivery: Lab turbidimeer Turb® 750 T; 4 batteries 1,5 V Typ AA; Wide range power supply; cable USB-A to USB-B mini; 6 empty vials, sticker for orientation marking; Cal.Kit Turb® 430/750 T; Cloth, quick guide, compact operation manual, CDROM with extended manual, CDROM with PC software Turb® Data; Inspection protocol
|