Your basket is empty
Home > Environmental > Sample Analysis > Turbidity > Turbidity meters Turb® 430 > Xylem - WTW Turb 430 T / Set 600326

Xylem - WTW Turb 430 T / Set 600326
Enlarge

Xylem - WTW Turb 430 T / Set 600326

PART NO: X-600326
Out of Stock (Ships in approx. 3 weeks)

Portable turbidimeter (90°, tungsten) acc. US EPA 180.1 incl. calibration standard kit (0.02-10-1000 NTU) in field case with table insert and accessories: Kal.Kit, 4 empty cuvette (28 m), cleaning tissues, screw driver, PC cable AK 540 B, instructions. Optional: LabStation LS Flex/430 with software package for GLP-compliant data export (order nr. 251 301) or rech.batt. Flex/430 (order nr. 251 300)

Turbidity meters Turb® 430
The turbidity meters Turb® 430 IR acc. to DIN ISO 27027 and Turb® 430 T acc. to US EPA 180.1 meet the requirements of the nephelometric measurement at 90° stray light for drinking water with highest precision, especially in the range < 1 NTU. The measuring range comprises 0-1100 NTU/FNU with automatic measuring range switching and is therefore suitable for water analysis as well as quality control, such as gasoline, cell cultures, as well as in the food and beverage industries.
The measurement and automatic 3-point calibration runs self-explanatory and menu-controlled. The calibration takes place via an AMCO Clear® standard set (0.02-10-1000 NTU). Up to 1000 data sets including sample ID numbers can be stored and later put out GLP-compliant via the PC software LSdata. The quality assurance of the measured results is supported by calibration interval setting with documentation.

Scope of delivery:
Turb® 430 IR / Turb® 430 T: Turbidity meter with calibration set 0.02-10.0-100 NTU
Turb® 430 IR/SET / Turb® 430 T/SET: Turbidity meter in case with table insert, calibration set 0.02-10.0-100 NTU, PC software LSdata, accessories
Specifications:
Measuring range: NTU 0 to 1100
FNU 0 to 1100
Resolution: 0.01 in the range 0.00 to 9.99
0.1 in the range 10 to 99.90
1 in the range 100 to 1100
Accuracy: ± 0.01 NTU or ± 2 % of measured value
Reproducibility: 0.01 NTU or < 0.5% vof measured value
Cell size: 28 x 60mm, 20 ml sample volume
Power supply: 4 x AA batteries for approx. 3000 measurements

Product Specification:
Product Type: Turb® 430 T/SET
Description: with tungsten lamp

Catalogue Number9.923 604
ManufacturerXylem - WTW
Manufacturer P/n600326
TypeTurb® 430 T/SET
PK1
Descriptionwith tungsten lamp
Search words