Your basket is empty

A. Kruss Optronic Gas Analyzer MAT1400
Enlarge

A. Kruss Optronic Gas Analyzer MAT1400

PART NO: 4AJ-6269836
Out of Stock (Ships in approx. 2 weeks)

with Zirconia dioxide sensor for oxygen

Gas analyzers MAT1400/MAT1500
Gas analyzers MAT1000 series are used for measuring oxygen concentrations (O2) or optional carbon dioxide concentrations (CO2).
Applies to all models of MAT1000 series:
-Easy and intuitive operation via touchscreen, even by untrained personnel
-Create your own method
-Available method parameters: scales, measurement procedures, limit values, comment and much more
-Selectable measurement methods: single or continuous
-Several single measurement procedures: manual, automatic, min/max detection
-Shortcut keys for most important functions
-Short measurement period, high accuracy even with small sample volume
-Validation possible
-Compliance with global standards such as GMP/GLP and 21 CFR Part 11
-High safety standard thanks to the traceability and repeatability of all measured data
-Documentation of all measured values of the last 999 measurements
-Integrated interfaces, flexible data export (USB, Ethernet, RS-232)
-Compact design in robust cast aluminum housing
-Built-in air pressure sensor
-Integrated flow rate monitoring
-IQ / OQ / PQ by A.KRÜSS Optronic or certified service partner
-On site service, maintenance, calibration and adjustment
Measurement of O2 and CO2 e.g. in modified atmosphere packaging, for work with modified atmosphere conditions.
-For high sample throughput
-For high O2 concentration
-High precision with the lowest O2 concentrations
-Wide measurement range
-Drift-free zirconia sensor (ZrO2)
-Low-drift non-dispersive infrared sensor (NDIR), model MAT1500
-Environmental compensated CO2 measurement, model MAT1500
Specifications:
O2 measuring range: 0 to 100 Vol.-%
Accuracy: 0.001 Vol.-% (<1 Vol.-%)
0.01 Vol.-% (<6.5 Vol.-%)
0.1 Vol.-% (<35 Vol.-%)
Resolution: 0.001 Vol.-%
CO2 measuring range (only MAT1500): 0 to 50 Vol.-%
Accuracy: ±0.5 Vol.-%
Resolution: 0.1 Vol.-%
Average measurement period: 5 to 10 sec
Heating time: 10 min
Interface: RS-232, USB, Ethernet
Dimensions (W x D x H): 200 x 350 x 150mm
Weight: 4.6 kg
Power consumption: 60 W
Power supply: 100-250 V, 50/60 Hz

Product Specification:
Product Type: MAT1400
Description: with zirconia sensor for oxygen

Catalogue Number6.269 836
ManufacturerA Kruss Optronic
Manufacturer P/nMAT 1400
Alternative RefMAT1400
TypeMAT1400
PK1
Descriptionwith zirconia sensor for oxygen
Search words

Gas analyzer(s)

Alternative Reference:MAT1400