Temperature (°C)
Range: Amb. +10 to 80
Fluctuation at 37°C: ±0.1
Variation at 37°C: ±0.5
Shaking System
Orbit Diameter (mm): 50.8
Speed Range, unstacked (rpm): 30 to 300
Speed Range, stacked (rpm): 30 to 200
Timer: 1min to 999hr 59min
Max. Load, unstacked (Kg): 20 at 300rpm
Max. Load, stacked (Kg): 15 at 200rpm and 20 at 150rpm
Dimensions (mm)
Platform; W x D: 755 x 520
Internal; W x D x H: 809 x 624 x 393
External; W x D x H: 1310 x 996 x 755
Weight (Kg): 375
Performance
• Temperature Range: Ambient +10°C to +80°C • Competitive high shaking speed up to 300rpm (stackable up to 200rpm) • Microprocessor PID control • Three point temperature calibration/automatic tuning • Wide speed range even with heavy workload. Ideal DD (Direct Drive) brushless motor and triple cam system support heavy load, quiet, maintenance-free operation and deliver high torque at high speeds and smooth rotation • Best effort run function intelligently manages its rpm to keep shaking even workload is out of its capacity. Automatic shaking speed adjustment in case of excessive workload such as unbalanced load placement, unusual vibrations caused by unstable floor or external shock • Pleasant test environment by the smooth acceleration and deceleration control. Smooth start and stop function prevents chemical spills from flasks or test tubes • High-velocity fan ensures uniform temperature distribution and rapid thermal recovery after the door is opened • Air-tight silicone door seal to provide excellent temperature uniformity
Convenience
• Intuitive touch-screen LCD controller with easy icons and logical menus • Easy-set digital timer for shaking operation (1min to 999hr 59min) • Wide sample monitoring. Clear observation of samples without affecting inner chamber’s environment thanks to the transparent viewing window with bright LED lamps • Repetitive tasks can be easily performed as the platform stops where it starts. Stopping the shaking platform always at the same position is highly beneficial for automated closing or sampling process • Built-in electrical outlet with a safety cover inside the chamber • Corrosion resistance stainless steel interior • Easy-access drain system with a quick disconnect valve offers great convenience to clean up spills • Built-in RS-232 port and USB port for external control and data collection • The unit comes with retractable foot caster, beneficial for easy mobility during installation or relocation
Safety
• Automatic run after power interruption • Temperature and shaking speed deviation alarm • Over-current protection • Triple Independent Temperature Monitor System: Temperature deviation alarm (high/low), electronic temperature limiter (high/low) and mechanical over-temperature limiter (high).
Specifications
Model
|
ISS-7200
|
ISS-7200R
|
Temperature
|
Range (°C / °F)¹
|
Amb. +10 to 80
/ Amb. +18 to 176
|
Amb. -20 (Min. 4) to 80
/Amb. -36 (Min. 7.2) to 176
|
Fluctuation at 37°C in flask
(±°C / °F)
|
0.1 / 0.18
|
Variation at 37° in flask
(±°C / °F)
|
0.6 / 1.08
|
0.5 / 0.9
|
Refrigerator (Hp)
|
-
|
1/6Hp
|
Shaking system
|
Motion type
|
Orbital
|
Amplitude size (mm/inch, dia.)
|
50.8 / 2
|
Speed range (rpm)
|
30 to 300 (stackable²):30 to 200
|
Speed stability²(rpm, %)
|
±1
|
Timer (count-down)
|
999 hr 59 min.
|
Max. Speed per permissible Load (rpm)
|
1 unit only 400 (max. 20 kg / 44.1 Ibs)
|
Stacked units 200 (max. 15 kg / 33.1 Ibs) 150 (max. 20 kg / 44.1 Ibs)
|
Dimension
|
Volume (L / cu ft)
|
198 / 70
|
Platform (W×D, mm / inch)
|
755×520 / 29.7x20.5
|
Interior (W×D×H, mm / inch)
|
809×624×393 / 31.9×24.6×15.5
|
Exterior (W×D×H, mm / inch)³
|
1 unit ; 1307 x 996 x 755 / 51.5 x 39.2 x 29.7 1 unit+stand-I or 2 units stacked ; 1307 x 996 x 1355 / 51.5 x 39.2 x 53.3 2 units+stand-II stacked ; 1307 x 996 x 1655 / 51.5 x 39.2 x 65.2 3 units stacked ; 1307 x 996 x 1955 / 51.5 x 39.2 x 77.0
|
Net weight (kg / Ibs)
|
300 / 661.4
|
320 / 705.5
|
1. The temperature control range of the ISS-7100 and ISS-7200 models is from 10°C above the ambient temperature of the equipment. (15°C to 40°C) to 80°C The temperature control range of the ISS-7100R and ISS-7200R models are from 20°C below the ambient temperature (18°C to 35°C) of the device to 80°C. (However, the lowest control temperature is 4°C)
2. Actual rpm allowable error for set rpm is 1rpm if setting is less than 100rpm, ±1% if setting is higher than 100rpm.
3. Product height: 600mm, height of Stand I: 600mm, height of Stand II: 300mm, height of wheel attachment: 155mm.
|